You are currently viewing Titration Endpoint vs. Equivalence Point: Reading a pH Curve

Titration Endpoint vs. Equivalence Point: Reading a pH Curve

Leer en español

As a drop of base falls into a beaker of citric acid, a pink swirl appears. With a pH probe, you can also watch for the steepest part of the pH curve. The colour change is how we spot the endpoint. The steep rise in pH helps us estimate the equivalence point.

What the pink colour tells you

In this recording from the UNED Costa Rica lab, 0.065 mol/L sodium hydroxide (NaOH) is added to a citric acid solution. The indicator is phenolphthalein. It starts out colourless, then turns pink as the pH enters its transition range.

The same beaker in two frames, with a clear solution on the left and a pink tint on the right.
The beaker before the drop falls (left) and as the pink colour appears (right). Footage from the UNED Costa Rica lab.

Watch what happens just after a drop lands. A streak of pink can disappear as the solution mixes. If your endpoint is the first faint pink that stays, give the stirrer time to mix the drop before you decide to stop.

How to estimate the equivalence point from a pH curve

At equivalence, the amount of base added matches the amount of acid according to the stoichiometry of the reaction. In this experiment, the pH jumps sharply over a small addition of base. To find that part of the curve, put the volume of NaOH added on the horizontal axis and the pH from the probe on the vertical axis.

Graph of pH against sodium hydroxide volume, with the steepest rise near 5.20 mL.
The recorded curve for citric acid titrated with 0.065 mol/L NaOH. The dashed line marks the estimated equivalence point.
Four readings near the steep rise in pH
NaOH added (mL)pH
5.1077.62
5.1437.92
5.1798.50
5.2149.23

In the recorded data, the largest change in pH per millilitre is between 5.179 and 5.214 mL. The midpoint gives an equivalence estimate of about 5.20 mL. That’s a reading from the curve. The pictures show the colour change itself.

What would you write down during the experiment?

Keep a record of both volumes. Write down one when the indicator meets your colour criterion and estimate the other from the curve. If they differ, go back to the moment you called the endpoint and check how closely spaced the pH readings are. The curve in this run puts equivalence above pH 7, so stopping at neutral pH would give you a different volume.

You can explore the acid–base titration lab or compare the two methods with the titration activity in LabsLand Teach.

This post is also available in Español.

Luis